TRAJECTORY CONTROL FOR A NEW TOTAL SKIN ELECTRON TREATMENT MACHINE
Olivier C L Haas*, Keith J Burnham*, John A Mills+
* Control Theory and Applications Centre (CTAC), Coventry University, Priory Street, Coventry CV1 5FB, UK, http://www.ctac.mis.coventry.ac.uk/
+ University Hospitals Coventry and Warwickshire NHS Trust, Coventry, UK
The paper focuses attention on the development of a control system for a new radiotherapy treatment machine that makes use of a betatron to deliver electron beams over the whole surface of a patient. The betatron moves above a patient at a fixed distance from the patient support system. This paper presents a method to predict the betatron trajectory, in terms of speed and position, according to variation in the patients geometry and is concerned with a simulation study of a PID control strategy implemented to control the speed of the betatron.
Keywords: conventional control, medical systems, medical applications, three term control, optimisation
Session slot T-Tu-M01: Modelling and Control for High Dependency Medicine/Area code 4c : Modelling and Control of Biomedical Systems

|